K700 High Precision Digital RTD Thermometer with Platinum Resistance Sensor Probe

High Precision Digital RTD Thermometers

High Precision digital RTD thermometer supplied with a stainless steel hand held probe.
Range -100°C ~ -300°C / -148°F ~ 572°F,
Resolution 0.1°C; 0.1°F.

Accuracy: at ( 23 ± 5°C )
Range                               Accuracy
-100°C ~ 300°C          ±(0.1% reading + 0.4°C)
-148°F ~ 572°F           ±(0.1% reading + 0.8°F)

Stainless steel hand held Probe PT100, length 120mm
Other Probes available on request

Product Code: M8018/K

Description

K700 High Precision Digital RTD Thermometers

Features
► Precision Platinum Temperature Sensor.
Suitable for PT100, PT500 and PT1000 sensors with a value of 3850
► High Accuracy With 0.01°C Resolution.
► Automatic Locked Push-pull Connector For Easy Probe Connection.
► Large Backlit LCD.
► Dual Display. °C & °F Scale Selectable
► MAX/MIN/AVERAGE Function.
► Data Hold Function.
► Relative Function.
► Auto Power Off.
► IP67 Waterproof
► Battery: 1.5V, AAA battery (3)
► Other Probes available on request

Product Code: M8018/K

Downloads

2023 Zeal General Catalogue page 69

More details

High Precision Digital RTD Thermometers

Specifications:

Numerical Display:                                                            4 digital liquid crystal display
Measurement Range:                                                        -100°C ~ 300°C; -148°F ~ 572°F
Resolution:                                                                          0.1°C; 0.1°F;
Sensor types:                                                                       Platinum resistance temperature sensor for pt-100, pt-500, pt-1000 (selectable) 4 wires.
ALPHA=0.003850
Measurement current:                                                      Approx 0.53mA
Maximum Voltage at Temperature couple Input:      60V DC, or 24Vrms AC

Environmental:

 Operating Temperature and Humidity:           0°C ~50°C (32°F ~ 122°F) ; 0 ~ 80% RH
 Storage Temperature and Humidity:               -10°C to 60°C (14°F ~ 140°F); 0 ~ 80% RH
 Altitude up to 2000 meters.

Temperature Coefficient:

For ambient temperatures from 0°C ~ 18°C and 28°C ~ 50°C, for each °C ambient below 18°C or above 28°C add the following tolerance into the accuracy spec.
0.01% of reading + 0.03°C ( 0.01% of reading + 0.06°F )